Micropilot S FMR531 with HART/4...20 mA
Operation
E Hauser
33
5.2.2
Display symbols
The following table describes the symbols that appear on the liquid crystal display:
Light emitting diods (LEDs):
There is a green and a red LED besides the Liquid Crystal Display.
Symbols
Meaning
ALARM_SYMBOL
This alarm symbol appears when the instrument is in an alarm state. If the symbol flashes, this indicates a
warning.
LOCK_SYMBOL
This lock symbol appears when the instrument is locked,i.e. if no input is possible.
COM_SYMBOL
This communication symbol appears when a data transmission via e.g. HART is in progress.
Calibration to regulatory standards disturbed
If the instrument is not locked or it cannot guarantee the calibration to regulatory standards, the situation
will be indicated on the display via the symbol.
LED (LED)
Meaning
red LED continuously on
Alarm
red LED flashes
Warning
red LED off
No alarm
green LED continuously on
Operation
Green LED flashes
Communication with external device