Micropilot S FMR531 with HART/4...20 mA
Wiring
E Hauser
27
Overvoltage protector
• The level transmitter Micropilot S is equipped with an internal overvoltage protector (600 Vrms
surge arrester) according to DIN EN 60079-14 or IEC 60060-1 (impulse current test 8/20
μ
s, Î
= 10 kA, 10 pulses). Additionally, the instrument is protected by a galvanic insulation of 500 Vrms
between the power supply and the (HART) current ouput. Connect the metallic housing of the
Micropilot S to the tank wall or screen directly with an electrically conductive lead to ensure
reliable potential matching.
• Installation with additional overvoltage protector HAW262Z/HAW56xZ (see XA081F-A "Safety
instructions for electrical apparatus certified for use in explosion-hazardous areas").
–
Connect the external overvoltage protector and the Micropilot S transmitter to the local
potential matching system.
– Potentials shall be equalised both inside and outside the explosion hazardous area.
– The cable connecting the overvoltage protector and the Micropilot S transmitter shall not
exceed 1 m in length;
– The cable shall be protected e.g. routed in an armoured hose.
Power supply
For stand alone operation recommended via two Hauser RN221N.
mm accuracy
For measurements with mm accuracy the measured variable must be transmitted using HART
protocol to ensure the necessary resolution.
4.2.1
Connection to Tank Side Monitor NRF590
See Page 25.
Summary of Contents for Micropilot S FMR531
Page 4: ...Brief overview Micropilot S FMR531 with HART 4 20 mA 4 Endress Hauser ...
Page 87: ...Micropilot S FMR531 with HART 4 20 mA Technical data Endress Hauser 87 ...
Page 98: ...Micropilot S FMR531 with HART 4 20 mA Index 98 Endress Hauser ...
Page 100: ...www endress com worldwide BA207F 00 en 04 09 71093754 CCS FM SGML 6 0 ProMoDo 71093754 ...