Micropilot S FMR531 with HART/4...20 mA
Trouble-shooting
E Hauser
73
9.3
Application errors
Error
Output
Possible cause
Remedy
A warning or alarm
has occurred.
Depending on the configuration
See table of
error messages
(see Page 71)
1.
See table of error messages
(see Page 71)
Measured value
(00) is incorrect
L00-FMR53xxx-19-00-00-en-007
Measured distance
(008) OK?
yes
→
1.
Check empty calibr. (005) and
full calibr (006).
2.
Check linearisation:
→
level/ullage (040)
→
max. scale (046)
→
diameter vessel (047)
→
Check table
3.
Check dip table
no
↓
Measurement in bypass
or stilling well?
yes
→
1.
Is bypass or stilling well
selected in tank shape?
2.
Is the pipe diameter
(007) correct?
3.
Pipe diameter correction (032 )
activated?
no
↓
Is "offset" (057) active?
yes
→
1.
offset (057) correctly set?
no
↓
An interference echo
may have been
evaluated.
yes
→
1.
Carry out tank mapping
→
basic setup
2.
Activate auto correction (031)
No change off
measured value on
filling/emptying
L00-FMR53xxx-19-00-00-en-006
Interference echo from
installations, nozzle or
extension on the
antenna
1.
Carry out tank mapping
→
basic setup
2.
If necessary, clean antenna
3.
If necessary, select better
mounting position (see Page 15)
20 mA/100%
F m/ft
E m/ft
4 mA/0%
t
→
D m/ft (008)
actual
expected
20 mA/100%
4 mA/0%
t
→
actual
expected
Summary of Contents for Micropilot S FMR531
Page 4: ...Brief overview Micropilot S FMR531 with HART 4 20 mA 4 Endress Hauser ...
Page 87: ...Micropilot S FMR531 with HART 4 20 mA Technical data Endress Hauser 87 ...
Page 98: ...Micropilot S FMR531 with HART 4 20 mA Index 98 Endress Hauser ...
Page 100: ...www endress com worldwide BA207F 00 en 04 09 71093754 CCS FM SGML 6 0 ProMoDo 71093754 ...