Micropilot S FMR531 with HART/4...20 mA
Commissioning
E Hauser
41
6.3
Basic Setup
L00-FMR531xx-19-00-00-en-002
SD
F
L
D
BD
E
flange:
reference point of
measurement
E
= empty calibr. (= zero)
settings in 005
settings in 006
D = distance (distance flange / product)
L
display in 0A5
in 0A6
F
= full calibr. (= span)
= level
display
SD = safety settings
BD = blocking dist.
settings in 015
settings in 059
...
(for bypass/stilling well)
(descr
iption see BA 217F)
measuring cond.
measuring cond.
pipe diameter
mapping
empty calibr.
full calibr.
basic setup (standard)
mounting
calibr
.
safety settings
linearisation
extended calibr.
option
first point corrects
further points correct linearity
offset
Commissioning
Summary of Contents for Micropilot S FMR531
Page 4: ...Brief overview Micropilot S FMR531 with HART 4 20 mA 4 Endress Hauser ...
Page 87: ...Micropilot S FMR531 with HART 4 20 mA Technical data Endress Hauser 87 ...
Page 98: ...Micropilot S FMR531 with HART 4 20 mA Index 98 Endress Hauser ...
Page 100: ...www endress com worldwide BA207F 00 en 04 09 71093754 CCS FM SGML 6 0 ProMoDo 71093754 ...