Technical data
Liquisys M CUM223/253
96
Hauser
11
Technical data
11.1
Input
Measured variables
Turbidity, suspended solids, temperature
Measuring range
Cable specification
Signal input
Digital communication
Temperature
measurement
NTC 30 k
at 25 °C (77 °F)
Binary inputs
Current input
4 to 20 mA, galvanically separated
Load: 260
at 20 mA (voltage drop 5.2 V)
CUS31:
0.000 to 9999 FNU/NTU
0.00 to 3000 ppm
0.0 to 3.0 g/l
0.0 to 200.0 %
CUS41:
0.00 to 9999 FNU/NTU
0.00 to 9999 ppm
0.0 to 300.0 g/l
0.0 to 200.0 %
Temperature:
-5.0 to +70.0 °C (+23 to +158 °F)
Cable length:
max. 200 m (656 ft.)
Voltage:
10 to 50 V
Power consumption:
max. 10 mA