Diagnostics and troubleshooting
Liquisys M CUM223/253
76
Hauser
E063
Below current output
range 1
Check measured value and current
assignment.
Yes
No
No
80
80
E064
Current output range 1
exceeded
Yes
No
No
80
80
E065
Below current output
range 2
Yes
No
No
80
80
E066
Current output range 2
exceeded
Yes
No
No
80
80
E067
Set point exceeded limit
contactor 1
Check configuration.
Yes
No
No
80
80
E068
Set point exceeded limit
contactor 2
Yes
No
No
80
80
E069
Set point exceeded limit
contactor 3
Yes
No
No
80
80
E070
Set point exceeded limit
contactor 4
Yes
No
No
80
80
E079
Measuring value outside
concentration table
Clean sensor, check table.
Yes
No
No
44
80
E080
Current output 1 range
too small
Decrease current output spreading
Yes
No
—
—
2
80
80
E081
Current output 2 range
too small
Yes
No
—
—
2
80
80
E084
Calibration warning
Calibration data are within limits but
deviate from standard values by a factor
of more than two.
No
No
No
80
80
E085
Incorrect setting for error
current
If the current range "0 to 20 mA" was
selected in field O311, the error current
"2.4 mA" may not be set.
No
No
—
—
2
80
80
E100
Current simulation active
No
No
—
—
2
80
80
E101
Service function active
Switch off service function or switch
device off and then on again.
No
No
—
—
2
80
80
E102
Manual mode active
No
No
—
—
2
80
80
E106
Download yes
Wait for download to finish.
No
No
—
—
2
80
80
E116
Download error
Repeat download.
Yes
No
—
—
2
0C
0C
E152
PCS alarm
Check sensor and connection.
Yes
No
No
44
44
E153
Offset
Adjustment range exceeded
No
No
No
80
80
E154
Below lower alarm
threshold for period
exceeding alarm delay
Perform manual comparison
measurement if necessary. Service
sensor and recalibrate.
Yes
No
No
-
3)
-
E155
Above upper alarm
threshold for period
exceeding alarm delay
Yes
No
No
-
-
E156
Current value undershoots
alarm threshold (CC
setpoint) for longer than
the set permissible
maximum period
Yes
No
No
-
-
E157
Current value exceeds
alarm threshold (CC
setpoint) for longer than
the set permissible
maximum period
Yes
No
No
-
-
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom.
cleaning start
PROFIBUS
status
Facty
User
Facty
User
Facty
User
PV
1)
Temp