Liquisys M CUM223/253
Commissioning
Hauser
55
R216
Enter alarm
threshold
9999 FNU
9999 NTU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m
200.0 %
If the alarm threshold is undershot/
overshot, this triggers an alarm with the
error message and error current at the
transmitter (note alarm delay in field F3).
If defined as a Min contact, the alarm
threshold must be < switch-off point.
If two factory settings are displayed the
one on the left side stands for CUS41 the
one on the right side stands for CUS31.
R217
Display status for
limit contactor
MAX
MIN
Display only.
R2 (2)
Configure limit
contactor for
temperature
measurement
LC PV = limit contactor
TU (1)
LC °C = limit contactor
T (2)
PID controller (3)
Timer (4)
Clean = Chemoclean (5)
By confirming with ENTER, another relay
function already switched on is switched
off and its settings are reset to the factory
settings.
R221
Switch function of
R2 (2) off or on
Off
On
Settings made for the limit contactor are
not deleted by switching the function off.
R222
Enter switch-on
temperature
100.0 °C
(212 °F)
-5.0 to 100.0 °C
(23 to 212 °F)
Never set the switch-on point and the
switch-off point to the same value!
R223
Enter switch-off
temperature
100.0 °C
(212 °F)
-5.0 to 100.0 °C
(23 to 212 °F)
Entering a switch-off point selects either a
Max contact (switch-off point < switch-on
point) or a Min contact (switch-off point >
switch-on point), thereby implementing a
hysteresis that is constantly required (see
"Illustration of the alarm and limit
functions" figure).
R224
Enter pick-up delay
0 s
0 to 2000 s
R225
Enter drop-out delay
0 s
0 to 2000 s
R226
Enter alarm
threshold (as
absolute value)
100.0 °C
(212 °F)
-5.0 to 100.0 °C
(23 to 212 °F)
If the alarm threshold is undershot/
overshot, this triggers an alarm with the
error message and error current at the
transmitter (note alarm delay in field F3).
If defined as a Min contact, the alarm
threshold must be < switch-off point.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
FNU
R216
9999
A.Thresh
R217
LC State
MAX
R2
Sel
.
Type
LC °C
R221
Function
off
R222
¡C
100.0
On value
R223
¡C
100.0
Off value
R224
s
On Delay
0
R225
s
Off Delay
0
R226
¡C
100.0
A.Thresh