Liquisys M CUM223/253
Commissioning
Hauser
69
C114 Calibration status is
displayed
o. k.
E. xxx
C115 Store calibration
results
yes
no
new
If C114 = E xxx, then only no or
new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
C1 (2)
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration data (6)
C121
Enter correct
concentration of the
third calibration
solution
Current value from C113
entire measuring range
If the calibration is performed with an
unknown sample concentration, but with
a definite dilution (1/10; 1/3;1), the
laboratory value is to be entered.
C122 Calibration status is
displayed
o. k.
Exxx
C123 Store calibration
results
yes
no
new
If C122 = E xxx, then only no or
new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
C1 (3)
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration data (6)
C131
Enter concentration
of the first
calibration solution
Current value from C111
entire measuring range
Coding
Field
Selection or range
(factory settings bold)
Display
Info
C114
o.k.
Status
E045
E084
E084
E045
20 %
50 %
200 %
500 %
C161 ... C163
Cancel Warning
Cancel
Warning
C115
yes
Store
C1
Corr
Calibrat
C121
FNU
1000.0
Concentr3
C122
Status
o
.
k
.
C123
yes
Store
C1
Edit
Calibrat
C131
FNU
100.0
Concentr1