Commissioning
Liquisys M CUM223/253
68
Hauser
Coding
Field
Selection or range
(factory settings bold)
Display
Info
C
Function group
CALIBRATION
Calibration settings.
C1 (1)
Select calibration
3-Pt = Three-point
calibration
(1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration data (6)
For data set 1 (B4), only the “Data”
function is accessible. The offset is reset
with 3 Pt and Edit.
Immerse sensor in the calibration solution (sample 1).
Immerse the sensor so that there is
sufficient distance to the tank wall (no
reflection).
C111
Enter concentration
of the first
calibration solution
Value from last calibration
Immerse sensor in the calibration solution (sample 2).
Immerse the sensor so that there is
sufficient distance to the tank wall (no
reflection).
C112
Enter concentration
of the second
calibration solution
Value from last calibration
C112
1.1 x C111
Immerse sensor in the calibration solution (sample 3).
Immerse the sensor so that there is
sufficient distance to the tank wall (no
reflection).
C113
Enter concentration
of the third
calibration solution
Value from last calibration
C113
1.1 x C112
C
CALIBRAT
C1
3-Pt
Calibrat
C111
FNU
100.0
Concentr1
C112
FNU
330.0
Concentr2
C113
FNU
1000.0
Concentr3