Liquisys M CUM223/253
Commissioning
Hauser
35
6.4.2
Setup 2 (Temperature)
In the SETUP 2 function group, you can change the temperature and wiper settings.
Coding
Field
Selection of range
(factory settings bold)
Display
Info
B
Function group
SETUP 2
Initial display in function group SETUP 2.
B1
Switch wiper
controller on or off
off
on
auto
If “auto”is selected, the wiper is operated
in combination with a cleaning function via
timer/Chemoclean (“wipe and clean”). In
this case B2 and B3 are not applicable.
B2
Enter operating
period of wiper
30 s
3 to 999 s
B3
Enter pause time
between two wiping
cycles
30 min
1 to 7200 min
B4
Select calibration
data set to be used
3
1 to 3
There are 3 calibration data sets stored in
every operation mode (A1). Data set 1
cannot be changed.
A hold is active during the load of a new
selected data set (independent of the
settings in S2).
B5
Copy data sets
no
1 -> 2
1 -> 3
2 -> 3
3 -> 2
Data set 1 cannot be changed (factory
setting). However, it can be used as a basis
for a customer calibration data set. To
work with the copy of a data set, select the
data set in field B4.
B6
Display measured
value with reflection
compensation?
yes
no
With CUS31 / CUS41: Displays the
measured value with or without reflection
compensation. Only effective in NTU, FNU,
ppm, mg/l.
B7
Enter corrected
temperature of the
temperature sensor
Current measured value
-5.0 to 100.0 °C
This entry can be used to calibrate the
temperature sensor to an external
measurement.
B8
Displays the
temperature
difference (offset)
Current offset
-5.0 to 5.0 °C
The difference between the temperatures
measured and entered is displayed.