Commissioning
Liquisys M CUM223/253
70
Hauser
C132
Enter concentration
of the second
calibration solution
Current value from C112
C132
1.1 x C131
C133
Enter concentration
of the third
calibration solution
Current value from C113
C133
1.1 x C132
C134
Calibration status is
displayed
o.k.
Exxx
C135
Store calibration
results?
yes
no
new
If C134 = E xxx, then only no or
new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
C1 (4)
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation
(4)
1-Pt = Single-point
calibration (5)
Data = Calibration data (6)
Only for solutions = 2 FNU / 5 ppm!
Backscatter from the immediate sensor
environment is compensated for clear
media.
C141
Enter correct
measured value
0.0 NTU
0.0 to 2.0 NTU
0.0 FNU
0.0 to 2.0 FNU
0.0 ppm
0.0 to 5.0 ppm
0.0 mg/l
0.0 to 5.0 mg/l
Only for the ranges NTU, FNU, ppm, mg/l
C142 Calibration status is
displayed
o.k.
Exxx
C143 Store calibration
results?
yes
no
new
If C142 = E xxx, then only no or
new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
Coding
Field
Selection or range
(factory settings bold)
Display
Info
C132
FNU
330.0
Concentr2
C133
FNU
1000.0
Concentr3
C134
o.k.
Status
C135
yes
Store
Refl
Calibrat
C1
C141
FNU
0.0
Real PV
C142
o.k.
Status
C143
yes
Store