Commissioning
Liquisys M CUM223/253
54
Hauser
Basic version does not include functions in
italic
.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
R
RELAY function
group
Relay contact settings.
R1
Select contact to be
configured
Rel1
Rel2
Rel3
Rel4
Rel3 (water) and Rel4 (cleaner) are only
available with the relevant version of the
transmitter.
If Chemoclean is used as the cleaning
method, Rel4 is not available.
R2 (1)
Configure limit
contactor for
turbidity
measurement
LC PV = limit contactor
TU
(1)
LC °C = limit contactor T (2)
PID controller (3)
Timer (4)
Clean = Chemoclean (5)
PV = process value
If Rel4 is selected in the R1 field,
Clean = Chemoclean cannot be selected.
By confirming with ENTER, another relay
function already switched on is switched
off and its settings are reset to the factory
settings.
R211
Switch function of
R2 (1) off or on
Off
On
All the settings are retained.
R212
Enter the switch-on
point of the contact
9999 FNU
9999 NTU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m
200.0 %
Never set the switch-on point and the
switch-off point to the same value!
(Only the operating mode selected in A1
is displayed.)
If two factory settings are displayed the
one on the left side stands for CUS41 the
one on the right side stands for CUS31.
R213
Enter the switch-off
point of the contact
9999 FNU
9999 NTU
9999 ppm / 3000 ppm
9999 mg/l / 3000 mg/l
300.0 g/l / 3 g/l
99.99 kg/l
99.99 t/m
200.0 %
Entering a switch-off point selects either a
Max contact (switch-off point < switch-on
point) or a Min contact (switch-off point >
switch-on point), thereby implementing a
hysteresis that is constantly required (see
"Illustration of the alarm and limit
functions" figure).
If two factory settings are displayed the
one on the left side stands for CUS41 the
one on the right side stands for CUS31.
R214
Enter pick-up delay
0 s
0 to 2000 s
R215
Enter drop-out delay
0 s
0 to 2000 s
R
RELAIS
R1
Sel
.
Relay
Rel1
R2
Sel
.
Type
LC PV
R211
Function
off
FNU
R212
9999
On value
FNU
R213
9999
Off value
s
R214
On Delay
0
s
R215
Off Delay
0