Commissioning
Liquisys M CUM223/253
32
Hauser
14. In B1, switch wiper controller on or off.
Press ENTER to confirm.
off
on
auto
15. In B2, enter operation period of wiper.
Press ENTER to confirm.
30 s
3 to 999 s
16. In B3, enter pause time between two wiping
cycles.
Press ENTER to confirm.
30 min
1 to 7200 min
17. In B4, select the calibration data set to be used.
Press ENTER to confirm.
3
1 to 3
18. In B5, select the data set to be copied.
Press ENTER to confirm.
no
1 -> 2
1 -> 3
2 -> 3
3 -> 2
19. In B6, select if the measured value shall be
displayed with reflection adjustment (for
solutions
2 FNU / 5 ppm).
Press ENTER to confirm.
yes
no
20. In B7, enter the corrected temperature of the
temperature sensor (for adjustment to external
measurement).
Press ENTER to confirm.
Current measured
value
-5.0 to 100.0 °C
21. In B8, the current offset is displayed.
Press ENTER to confirm.
Current offset
-5.0 to 5.0 °C
22. In B9, enter gas bubble barrier. For clear media
with measured values <1000 NTU the gas bubble
barrier has to be set to 100 %.
Press ENTER to confirm.
3.0 %
0.1 to 100 %
User input
Setting range
(Factory settings,
bold)
Display
B1
Off
wiper
B2
s
30
CleanTime
B3
min
30
PauseTime
B4
3
data set
B5
no
copy data
B6
yes
refl.adj.
B7
°
C
0.0
RealTemp
B8
°
C
0.0
Temp.Offs
B9
%
3.0
GasBubble