Commissioning
Liquisys M CUM223/253
42
Hauser
Basic version does not include functions in
italic
.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
O
CURRENT OUTPUT
function group
Configuration of the current output (does
not apply for PROFIBUS).
O1
Select current output
Out1
Out 2
Output 2 not available for all versions.
A characteristic can be selected for every
output.
O2
Select measured
variable for 2nd
current output
°C
mg/l
Contr
R237 = curr (current output 2) can only
be selected if O2 = Contr is selected (relay
board required).
O3 (1)
Enter or output
linear characteristic
lin = linear
(1)
sim = simulation (2)
tab = table (3)
The characteristic can have a positive or
negative slope for the measured value
output.
In the case of actuating variable output
(O2 = Contr), an increasing current
corresponds to an increasing actuating
variable.
O311 Select current range
4 to 20 mA
0 to 20 mA
O312
0/4 mA value:
Enter corresponding
turbidity or
temperature value
0.000 FNU
0.000 NTU
0.000 ppm
0.000 mg/l
0.000 g/l
0.000 kg/l
0.000 t/m
0.000 %
0.000 °C
Here you can enter the turbidity or
temperature value at which the min.
current value (0/4 mA) is applied at the
transmitter output.
Minimum distance between 0/4 mA and
20 mA value: see field O313
Display format from A3
O313
20 mA value:
Enter corresponding
turbidity or
temperature value
10.00 FNU
10.00 NTU
10.00 ppm
10.00 mg/l
300 g/l / 3.00 g/l
99.99 kg/l
99.99 t/m
10.0 %
100 °C
Here you can enter the turbidity or
temperature value at which the max.
current value (20 mA) is applied at the
transmitter output.
Display format from A3
If two factory settings are displayed the
one on the left side stands for CUS41 the
one on the right side stands for CUS31.
O3 (2)
Simulate current
output
Lin = linear (1)
Sim = simulation
(2)
Tab = table (3)
Simulation is not ended until (1) or (3) is
selected.
O
OUTPUT
O1
Sel. Out
Out1
O2
Sel. Out2
¡C
lin
O3
Sel.Type
O311
Sel
.
Range
4
-
20
O312
FNU
0/4 mA
0.000
O313
20 mA
10.00
FNU
20 mA
10.00
O3
sim
Sel.Type