Liquisys M CUM223/253
Commissioning
Hauser
71
C1 (5)
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration data (6)
For FNU:
Adaption C164, C165
For ppm, mg/l:
up to 500 - adaption C164, C165
above 500 - adaption C166
For g/l, %:
Adaption C166. An existing basic
calibration (three-point) is corrected by
the single-point calibration.
C151
Enter current
calibration value
Current measured value
Entire measuring range
C152 Calibration status is
displayed
o.k.
Exxx
C153 Store calibration
results?
yes
no
new
If C152 = E xxx, then only no or
new
(Exception: calibration warning E84). If
new, return to C. If yes / no, return to
“Measurement”.
C1 (6)
Select calibration
3-Pt = Three-point
calibration (1)
Corr = Three-point
correction (2)
Edit = Edit calibration (3)
Refl = Fitting with
reflection compensation (4)
1-Pt = Single-point
calibration (5)
Data = Calibration data
(6)
C161 Calibration point 1 is
displayed
Comparison value
Deviation relative to standard sensor
(= 100 %)
C162 Calibration point 2 is
displayed
Comparison value
Deviation relative to standard sensor
(= 100 %)
C163 Calibration point 3 is
displayed
Comparison value
Deviation relative to standard sensor
(= 100 %)
Coding
Field
Selection or range
(factory settings bold)
Display
Info
C1
1-Pt
Calibrat
C151
FNU
FNU
Real PV
C152
o.k.
Status
C153
yes
Store
C1
Data
Calibrat
C161
%
101.4
Concentr1
C162
%
99.3
Concentr2
C163
%
98.7
Concentr3