Liquisys M CUM223/253
Appendix
Hauser
105
a0007368
Field for customer
settings
O335
O336
O334
Entry of x value
(meas. value)
0.000 NTU
FNU / g/l / %
Z7
Entry of y value
(current value)
0.0 ... 20.0 mA
4.00 mA
Table status o.k.
no
yes;
Feedforward control = 1
at
0 … 100%
50%
R225
R226
R235
R236
R237
R239
R238
R215
R216
R255
R256
R257
R259
R258
R21
7
Display of LC status
MIN
MAX
Display of LC status
MIN
MAX
R22
7
F8
F
7
F9
P
7
P8
Enter setpoint
0 ... 9999 NTU
0.000 NTU
Automatic start of
cleaning function
yes
(not always displayed,
see error messages)
no;
Activate error current
for previously set
error
yes
no;
Select “next error”
or return to menu
R
next=next error
¬
Set max. perm. period
for lower limit exceeded
0 ... 2000 min
60 min
Entry of post-rinse time
0 ... 999 s
20 s
Number of repeat cycles
0 ... 5
0
Set interval between
two cleaning cycles
(pause time)
1 ... 7200 min
360 min
Number of cleaning
cycles without
cleaning agents
0 ... 9
0
Set minimum
pause time
1 ... R357 min
120 min
Entry of derivate
action time Tv
0.0...999.9 min
0.0 min
Selection of
controller characteristic
inv = inverted
dir = direct
Selection
freq = pulse frequency
curr = current output 2
Ien = pulse length
Entry of max.
pulse frequency
60 ... 180 1/min
120 1/min
Entry of
pulse interval
0.5...999.9 s
10.0 s
Dropout delay
setting
0 ... 2000 s
0 s
Setting of alarm threshold
(as an absolute value)
100.0 °C
–5.0 ... +100.0 °C
Dropout delay
setting
0 ... 2000 s
0 s
Setting of alarm threshold
(as an absolute value)
entire measuring range
9999 FNU; 9999 ppm (mg/l);
300.0 g/l; 200.0 %
R2311
R2310
Entry of
basic load
0 ... 40 %
0 %
Entry of minimum
switch-on time t
0.1...5.0 s
ON
0.3
s