ASP station 2000
6 Commissioning
Hauser
93
6.6
Configuration using Profibus
For a detailed description of the configuration of the ASP station 2000 using Profibus-
DP, please refer to the "Appendix to the operating manual ASP station 2000: DP-Slave-
Modul_is Pro Gate“ (see 10.10 “Documentation”).
Creating programmes - Switching programmes
Switch
Set
↵
Creating programmes
↵
Switch 1,2
Sample:
sampling type in the switching programme.
Distribution:
distribution mode for the switching programme.
-
Sample
Set
↵
Creating programmes
↵
Switch 1,2
↵
Sample
Time:
sampling acc. to time.
Volume:
sampling acc. to volume (connect volume signal!).
Ext. sig.:
sampling on external signal (connect signal input!).
Flow (optional):
sampling acc. to flow (connect volume
signal).
Time
Distribution
Set
↵
Creating programmes
↵
Switch 1,2
↵
Distribution
Time:
bottle change after a specific time.
Number:
bottle change after a specific number of samples.
Ext. sig.:
Bottle change on external signal (connect signal input!).
Time
Parallel sample
Set
↵
Creating programmes
↵
Switch 1,2
↵
Parallel sample
Activation and configuration of a sample in separate bottles, in
parallel to the switching programme.
No
Creating programmes - Event programme
Event prog.
Set
↵
Creating programmes
↵
Event prog.
Sample:
sampling type in the event programme.
Distribution:
distribution mode for the event programme.
-
Sample
Set
↵
Creating programmes
↵
Event prog.
↵
Sample
↵
mode
Time:
Sampling acc. to time
Quantity:
Sampling acc. to volume (connect volume signal!).
Ext. sig.:
Sampling on external signal (connect signal input!).
Flow (optional):
Sampling acc. to flow (connect volume signal).
Once:
Single sample.
Time
Distribution
Set
↵
Creating programmes
↵
Event prog.
↵
Distribution
↵
mode
Time:
Bottle change after a specific time.
Number:
Bottle change after a specific number of samples.
Ext. sig.:
Bottle change on external signal (connect signal input!).
Time
Switch
Set
↵
Creating programmes
↵
Event prog.
↵
Distribution
↵
Switch
Kein Einf:
No bottle change for event sample.
Change:
Change to the next empty bottle.
1-x bottles:
Between 1 and x bottles are filled during event sampling.
Kein Einf
Service
Service
Set
↵
Service
General:
Software name, software version, software option, CPU
number, preset, view angle setting (contrast)
Run times:
Test:
Service code necessary!!
Calibration:
Tap calibration (flow-proportional option: sample volume
calibration)
-