ASP station 2000
6 Commissioning
Hauser
81
The internal datalogger is activated and configured as follows:
6.4.3
Configuring the outputs
Configuring the outputs
Menu structure:
Fig. 27:
Outputs menu structure
Fig. 26:
Activating the internal datalogger
1.
Under <SET -> Basic Settings ->, select the item
“mem. interval”.
2.
Set the scan rate for the analogue signal. The
capacity of the memory is then displayed
automatically in the display. (Example scan rate =
10 sec. -> memory capacity = 170 hours). At a scan
rate of 0 seconds (default setting), the analogue
signal is not recorded. Pressing the <AUT key>
activates the internal datalogger.
!
Note!
Pressing the <AUT key> irreversibly deletes all the data
saved to the internal datalogger up to that point and
starts recording of the new data.