6 Commissioning
ASP station 2000
88
EHauser
Sample mode
Menu structure:
Fig. 37:
Sample mode menu structure
The event programme contains the same sampling options (time, quantity, ext. signal
and flow) as those of the main and switching programmes. In addition, you can select
the function <ONCE>. In this function, the sampler takes a sample in the event
programme once and then returns immediately to the main programme if the event
signal is no longer present.
Distribution
You can reserve separate bottles for the event programme. With the exception of
parallel reserve sampling, the following applies when distributing to bottles:
The first bottle group of a distribution is reserved for the main programmes.
The second bottle group is reserved for the switching programmes.
The last bottle group is reserved for the event programme.
Menu structure:
Fig. 38:
Distribution menu structure
You can define the bottles for the event programme as follows:
no change
There is no bottle change when the system switches to the event
programme.
move
The next empty bottle is filled when the system changes to the event
programme.
1-9 bottles
When the system switches to the event programme, 1-9 bottles are filled from
the last distribution bottle group. The number of bottles you can reserve for
the event programme is dependent on the total number of bottles (max.
9 bottles).
event program
Sample
Sample >
time
Distribution >
-mode
: time
quantity
Parallel sample >
time
:
00:15 ext. sig.
volume
:
100ml flow
shots
:
01 once
Esc
â
<-'
Esc
â
á
<-'
event program
Distribution
Sample >
time
Distribution >
-mode
: time
quantity
Parallel sample >
time
:
01:00 ext. sig.
switch
: no change
Esc
â
á
<-'
Esc
â
á
<-'