6 Commissioning
ASP station 2000
86
EHauser
•
Programme end: the unit stops sampling automatically when it has run through the set
programme.
•
No: the unit runs through the programme in an infinite loop. Do not forget to empty the
bottles.
•
Time: the sampling programme stops at a settable time.
You can choose between continuous operation and operation at different intervals for
the operating mode.
•
Day: operating time at two settable times per day.
•
Week: operating time on three settable days per week.
•
Interval: operating at specific time intervals.
Synchronisation
Menu structure:
Fig. 33:
Synchronisation menu structure
You can use synchronisation to assign specific filling times to specific bottles. In this
way you could fill bottle 1 between 00:00 and 02:00 and bottle 2 between 02:00 and
04:00 etc. For this, you have the following options:
•
AUT-button: the sampling and bottle change times are not synchronised.
•
Synchronisation TIME: sampling starts with the first bottle. The change to the next
bottle is synchronised. Example: a time of 2 hours was set for the bottle change and
the point 00:00 was set for the synchronisation. If the programme is started at 05.23,
bottle 1 is filled. The system then changes to bottle 2 at 06.00, to bottle 3 at 08.00 etc.
•
Synchronisation TIME+BOT.: each bottle is allocated a specific filling time. Example:
00:00-02:00: bottle 1; 02:00-04:00: bottle 2; 04:00-06:00: bottle 3, etc. For example is
you start the programme at 10:00, the unit will start filling bottle 6.
6.4.6
Creating switching programmes
Creating switching
programmes
Sample mode
Menu structure:
Fig. 34:
Sample mode menu structure
In the switching programmes, like in the main programmes, you can trigger sampling
by time, quantity, flow or an external signal.
Program 1
Name
Program 1
Sample >
Synchronisation
Distribution >
aut-button
Start-Stop >
-mode
: time
time
Synchronisation >
time+bot.
Parallel sample >
Esc
â
á
<-'
Esc
â
á
<-'
switch 1
Sample
Sample >
time
Distribution >
-mode
: time
quantity
Parallel sample >
time
:
00:15 ext. sig.
volume
:
100ml flow
shots
:
01 once
Esc
â
<-'
Esc
â
á
<-'