6 Commissioning
Hauser
91
Sampling
Set
↵
Basic settings
↵
Sampling
Time set.:
You can change the rinsing, suction, dosing and emptying
times manually.
Delay:
You can delay the start of sampling (e.g. after an external
signal) by up to 99 seconds.
CS 1/2:
You can adapt the sensitivity of conductivity probes to different
media (e.g. high for media with poor conductivity).
Dosing:
Support dosing operation with pressure (e.g. for low suction
heights) or make unpressurised.
No flow:
Count and save if there is no inflow during sampling (e.g. with
blockages).
Automatic
00 sec.
Standard
No press.
Count
Inputs
Set
↵
Basic settings
↵
Inputs
Function for digital input 1, 2 and 3 and for setting the
analogue input.
No
Outputs
Set
↵
Basic settings
↵
Outputs
Function for relay outputs 1, 2 and 3.
Power down
Memory interval
Set
↵
Basic settings
↵
Memory interval
Time interval for internal recording of measured flow values. 0
sec = no recording.
0000 sec
Thermostat
Set
↵
Basic settings
↵
Thermostat
Configuration of thermostat in sample compartment; setting of
sample compartment temperature and defrost times.
04 °C
Interfaces
Set
↵
Basic settings
↵
Interfaces
Transmission rate and definition of the RS232 interface
Creating programmes
Number
Set
↵
Creating programmes
↵
Number
1:
1 main programme active
2:
1 main programme and 1 switching programme active
1+Event:
1 main programme and one event programme active
2+Event:
1 main programme, 1 switching programme and 1 event
programme active
1
Name
Set
↵
Creating programmes
↵
Name
Enter names for the four main programmes, two switching
programmes and the event programme.
programme 1-4;
1-U1; 1-E;
1-U1-E
Switch (only for an
active switching
programme)
Set
↵
Creating programmes
↵
Switch
Day:
Switch between main and switching programmes on two
specific times a day.
Week:
Switch between main and switching programmes on two
specific weekdays.
Q greater:
Switch when flow greater than limit value.
Q smaller:
Switch when flow smaller than limit value.
External:
Switch on external signal.
Day