multi X 2500
Maintenance and care
89
6.
Unscrew the particle filters on the inside
using a 5-mm Allen key.
7.
Insert and tighten the new particle
filter.
8.
Screw in the gas connections and tighten them with an open-ended wrench.
9.
Connect the gas hoses. Make sure to attach the gas hoses to the correct connection
terminal.
10.
Reconnect the chlorine module and the sample introduction module.
11.
Open the gas supply.
12.
Plug in the mains cables to the connections on the multi X 2500 and the chlorine
module and switch on the devices using the mains switch.
The multi X 2500 is ready for operation again.
10.3
Maintenance activities (standard version)
10.3.1
Removing and cleaning the open combustion tube
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear protective clothing
when working at the sulfuric acid container. Observe all specifications in the safety
data sheet!
CAUTION
Risk of burns! Only remove the combustion tube when the device is cold!
Before switching off set the furnace temperature in multiWin to 20 °C and exit
multiWin. Otherwise there is a risk of burns when checking the system for leaks after
installation!