Safety instructions
multi X 2500
10
2.2
Requirements for the operating personnel
The multi X 2500 analyzer must only be operated by qualified specialist personnel
instructed in the use of the analyzer. The instruction must also include conveying the
content of this user manual and the user manuals of other system components or add-
on equipment.
In addition to the safety at work instructions in this user manual the generally
applicable safety and accident prevention regulations of the respective country of
operation must be observed and complied with. The operator must ascertain the latest
version of these regulations.
The user manual must be accessible to the operating and service personnel at any
time!
2.3
Safety instructions, transport and commissioning
The analyzer is always installed by the customer service department of Analytik Jena or
its authorized and trained specialist personnel. Independent assembly and installation
are not permitted. Incorrect installation can create serious hazards.
Observe the following notices during transport:
Insufficiently secured components pose a risk of injury! During transport the
components of the equipment must be secured in accordance with the instructions
in the user manual. The combustion furnace, auto-protection assembly, all
components made of glass and the measuring cell must be removed before
transporting the device.
Only transport the analyzer in its original packaging! Ensure that the transport
protections have been fitted and the analyzer is completely empty.
To prevent health damage the following must be observed when moving the
analyzer in the laboratory (lifting and carrying):
The weight of the multi X 2500 is 25 kg. For reasons of safety 2 persons are
required to transport the analyzer and must position themselves on both sides of
the equipment.
Because the analyzer does not have carrying handles, the device must be gripped
firmly with both hands at the lower end. The doors must be closed.
Risk of damage to health due to improper decontamination! Perform a professional
and documented decontamination of the device before returning it to Analytik
Jena. The decontamination report is available from the customer service
department when registering the return. Analytik Jena must refuse acceptance of
contaminated devices. The sender may be liable for damage caused by inadequate
decontamination of the device.