Transport and storage
multi X 2500
114
12
Transport and storage
Observe the safety instructions in chapter "Safety instructions, transport and
commissioning", p. 10. Take particular care when transporting the basic device and the
chlorine module to prevent damage from impact or vibration.
Use the original packaging for transporting the basic device and its system components
(e.g. electrodes). Use individual, fracture-proof packaging for all glass parts.
WARNING
Concentrated sulfuric acid can cause severe chemical burns! Wear appropriate
protective clothing when working at the sulfuric acid container. Observe all
specifications in the safety data sheet!
12.1
Prepare device for transport (standard version)
1.
Exit the multiWin control software and allow the analyzer to cool down.
2.
Use the device switch to switch off the basic device and the chlorine module
respectively.
3.
Cut the gas supply.
4.
Disconnect the mains plugs from the power outlets.
5.
Detach all connections on the rear panel of the basic device and the rear panel of
the chlorine module:
Gas connectors
Serial connection cable
Mains cable
Exhaust air hose
1.
Detach the hose connection to the measuring cell and the connection to the
combustion tube (forked clamps) on the sulfuric acid container (vertical).
2.
Empty and rinse the sulfuric acid container. Remove the collection tray of the
combustion furnace.
3.
Remove the combustion tube and the oxygen sluice from the combustion furnace.
4.
Remove the combustion furnace (
→
"Removing the combustion furnace", p. 116).
5.
Detach all hoses from the basic device or put the hose ends into small bags and
attach them to the device with adhesive tape.
6.
Secure the top cover of the basic device and the maintenance flaps on the right-
hand side of the device with adhesive tape.
In basic device