AOX determination
multi X 2500
66
1.
Exit the multiWin software. Allow the multi X 2500 to cool down. Then, use the
mains switch to switch off the basic device and the chlorine module and cut the gas
supply.
2.
Tilt the combustion furnace into horizontal position.
3.
Hold the auto-protection assembly with the left hand and
pull the knob of the clamping mount with the right hand to
open the lock.
Attach the auto-protection assembly to the combustion
furnace.
4.
Screw hose no. 8 (1) into the connection.
5.
Plug in hose no. 11 (2). Push down the ring of the plug-in
connector to do this.
6.
Screw the gas transfer line (3) into the connection at the
side.
7.
Insert the electrical connections for the auto-protection
assembly (4) and for the gas transfer line (5) into the
terminals.
8.
Route the gas transfer line through the opening at the top
left of the basic device into the chlorine module.
9.
Open the pneumatic seal on the auto-protection assembly.
Move the toggle switch (6) up so that the combustion tube
can be inserted.