AOX determination
multi X 2500
58
Preparing the chlorine module:
1.
Fill the adsorption tube (1) with activated carbon and insert
it into the chlorine module (
→
in the adsorption tube", p. 92).
2.
Insert the measuring cell "sensitive" with lid (2) into the
chlorine module.
3.
Insert the combined electrode into the opening (3) of the
measuring cell.
4.
Insert the storage container for the combined electrode (4)
into the bracket in the door of the chlorine module.
5.
Connect the gas outlet of the measuring cell (5) via the olive-
tip suction adapter and hose no. 5 to the upper end of the
adsorption tube.
6.
Connect the lower end of the adsorption tube to the intake
pump using hose no. 4.
7.
Attach the measuring gas hose to the connection of the
combined electrode (6).
8.
Electrically connect the combined electrode and the
measuring cell to the inside of the rear panel of the chlorine
module.
Connection combined electrode (1)
Connection measuring cell (2)