POX determination
multi X 2500
86
Fig. 46 Connection of the argon hose to the basic device and the POX container
13.
Create a new method under M
ETHOD
M
ETHOD NEW
and release it.
14.
Defining the analysis parameters: sample volume, furnace temperature, POX
purging time, dosing time, delay time, integration time.
The measurement time is the sum of all the configured times.
15.
Activate the new method or an already existing method using the menu command
M
ETHOD
M
ETHOD
–
ACTIVATE
.
16.
Starting the measurement:
Select S
TART
S
TART
-A
NALYSIS
.
Select an analysis group or create a new one.
Create an analysis sequence.
Enter the sample ID for all samples in the field N
AME
.
Confirm the entry with [OK].
Click the [S
TART MEASUREMENT
] button.
The prepared analysis sequence is processed.
Follow the prompts in the multiWin control and analysis software, attach the purging
equipment to the sample container and confirm with [OK].