Maintenance and care
multi X 2500
94
10.4.2
Checking the system for tightness
NOTICE
Risk of destroying the internal MFM by corrosive gases! Only use the flow monitoring
set included in the scope of delivery to check the system tightness. Make sure that the
sulfuric acid container is completely empty.
The system tightness is not tested automatically. The flow monitoring set delivered
with the device allows the user to test the system's tightness manually. The test
establishes whether the supplied gas flow corresponds to the target flow rate. This
allows detecting leaks and obstructions.
Fig. 47 Flow monitoring set
1.
Open the front doors of the analyzer and the chlorine module.
2.
Drain the entire sulfuric acid from the sulfuric acid container with outlet in the
chlorine module. Clean and dry the sulfuric acid container (
→
sulfuric acid and cleaning the sulfuric acid container with outlet", p. 101).
NOTICE! Always clean and dry the sulfuric acid container. Check the hose system for
deposits of acid in form of small drips. If required, clean the hose system. Even
minor traces of sulfuric acid in the gas path will damage the MFM.
3.
Disconnect the measuring gas hose from the gas inlet tube (measuring cell "high
sensitive") or from the combined electrode (measuring cell "sensitive" / "high
concentration").
4.
Remove the "MFM in" screw connection above the cover of the control electronics
in the analyzer.
5.
Use the flow monitoring set to link the "MFM in" connection (1) with the
measuring gas hose (2) in the chlorine module.