Fault removal
multi X 2500
110
222
Sampler: wrong syringe size
Cause
Remedy
No syringe inserted
Insert syringe into autoX 112
Dosing volume in the method greater than inserted
syringe
Adjust dosing volume
The method for dosing liquids is to be enabled and the
gripper and solids track are still inserted
Insert sample rack for liquid samples into
autoX 112
Initialize
224
Sampler: wrong gripper
Cause
Remedy
No gripper inserted
Insert gripper into autoX 112
The method for solids is to be enabled and the syringe
and sample rack for liquid samples are still inserted
Insert solids rack into autoX 112
Initialize
225
Sampler: position not reached
Cause
Remedy
autoX 36 – target position is not reached
Check autosampler for mechanical hindrance,
remove hindrance if required
Replace rack
Initialize
226
Sampler: runtime exceeded
Cause
Remedy
Message from autoX 112 sampler saying that the
movement is completed takes too long (autoX 112
defective)
Record log files
Report to customer service
230
ABD: no connection
Cause
Remedy
Communication error after ABD was detected during
initialization
Acknowledge message
Initialize
232
Flame sensor error
Cause
Remedy
Flame sensor calibration failed
Record log files
Report to customer service
260
Sample handling missing
Cause
Remedy
No sampler module detected
Connect at least one sample introduction module
(autoX, auto-injector, ABD, MBD)
Initialize
270
Autoinjector - syringe: no connection
Cause
Remedy
No communication with auto-injector
Acknowledge message
Initialize
271
Autoinjector - syringe: runtime exceeded
Cause
Remedy
Communication error with auto-injector
Completion message at dosing end not received
Acknowledge message
Check auto-injector
Initialize