multi X 2500
Fault removal
109
200
Gasbox: no connection
Cause
Remedy
Communication error
Acknowledge message
Initialize
201
Gasbox: error while setting up nominal flow
Cause
Remedy
Communication error
Acknowledge message
Initialize
202
203
204
205
Gasbox: conversion error 1
Gasbox: conversion error 2
Gasbox: conversion error 3
Gasbox: conversion error 4
Cause
Remedy
Communication error
Acknowledge message
Initialize
206
Gas pressure error
Cause
Remedy
Overpressure in the analysis system due to obstructed
gas paths, e.g.:
combustion tube (inlet, glass drip)
filter in the auto-protection assembly
Kinked gas hoses
Connectors/couplings clogged
Gas inlet to measuring cell
Adsorption tube
See also user manuals of the system modules.
WARNING! Greatest caution is necessary during an
overpressure in the system! Never switch off a device
that is subject to overpressure!
Do not introduce any samples
Do not close multiWin
Do not switch off basic device and module
Do not switch off the gas supply
Manually ventilate device via chlorine module:
Manual ventilation:
WARNING! Concentrated sulfuric acid! Wear safety
clothing.
Detach hose connection to measuring cell.
Detach gas transfer line from sulfuric acid
container with outlet and drain sulfuric acid.
Detach gas transfer line from auto-protection
assembly.
After the overpressure has dissipated:
Switch off devices, cut gas supply.
Search for cause for overpressure and correct the
defect.
220
Sampler: no connection
Cause
Remedy
Communication faulty after sampler was detected
during initialization
Acknowledge message
Initialize
222
Sample boat: broken
Cause
Remedy
Boat broken during removal from the combustion tube
(only if ABD is used)
Remove broken boat from the system
Initialize