multi EA 5000
Sulfur analysis with the UVFD 5000, MPO 5000
71
1
2
3
4
5
6
7
8
S-UVF
Figure 44Rear of the sulfur detector
1 Chemical ozone decomposer (MPO)
2 Measuring gas inlet
3 Measuring gas outlet
4 Service interface and programming
button.
5 Interface to the basic module
6 Power connection
7 Fuse holder
8 Equipment switch
The diagram on the rear explains the connector allocation.
9.2
Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡
Always connect the modules to power when they are switched off.
}
Place the detection module to the left of the basic module. In the event of a series of
detection modules: place the detection module to the left or right of the others.
}
Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
}
Connect the detection module with the basic module via the interface:
S-UVF interface on the rear of the detection module
S-UVF interface on the rear of the basic module
}
Connect the measuring gas hose of the basic module to the "sample in" gas inlet on
the rear of the module.