Nitrogen analysis with the CLD 5000
multi EA 5000
48
7.3
Operation
CAUTION
Risk of respiratory problems due to leaking ozone
If the gas hoses have not been properly connected to the ozone generator, ozone may
leak out of the detection module.
¡
If you can smell ozone, switch off the device and check the connection of the gas
hoses on the ozone generator.
}
Switch on the basic module and the detection module.
ü
The devices boot up. The status LED on the front of the basic module light up in
green after approx. 30 s.
ü
The LED on the front of the detection module flashes during the run-in period.
Depending on the detector, the run-in period can take up to 30 min. After this,
the LED will light up continuously. Starting a measurement is only then possible.
}
Open the gas supply and set the required gas pressure.
}
Switch on the PC.
}
Start the control and analysis software and login with your username and password.
}
Initialize the analysis system by clicking on [Initialize analyzer].
ü
The initialization and automatic detection of all connected components will be
carried out.
}
Ready the samples.
}
Activate a pre-existing method via the Method |Method - activate menu item.
}
Alternatively: Create a new method in the Method |Method - new menu. Select the
measurement parameter in the method. Release and activate the method.
}
Select Start |Start - Analysis in the menu.
}
Select an analysis group or create a new one and confirm via [OK].
}
Create an analysis sequence.
}
Enter the sample ID for all sample in the Name field.
}
Release all sequence lines
}
Confirm the entries with [OK].
}
Click the [Start Measurement] button.
ü
The prepared analysis sequence is processed.
For manual sampling, follow the instructions in the software.