multi EA 5000
Maintenance and care
149
1
2
3
Figure 69Absorber
1 Hose 6 connection
2 Holding clamp
3 Absorber
14.10.3 Replacing the chemical ozone decomposer
WARNING
Risk of electric shock
High voltages are present in the interior of the device, which can lead to electric shock if
contacted.
¡
Before opening: Switch off the device via the power switch.
¡
Disconnect the power cable from the socket.
CAUTION
Risk of burns from the thermal ozone decomposer!
¡
Only perform maintenance in the inside of the device when cold/allow the device to
cool sufficiently.
ð
Replace the entire chemical ozone decomposer annually. Optionally, have the de-
composer replaced by the customer service department.