multi EA 5000
Chlorine analysis with the Cl module
61
Installing the detection module
1
2
3
4
5
6
7
3
5
4
Figure 41Connection of the heated gas transfer lines to the sulfuric acid container
1 Safety attachment
2 Sulfuric acid container
3 Hose for measuring gas infeed
4 Connector
5 Banjo bolt
6 Heated gas transfer line
7 Measuring gas transfer to the mea-
suring cell with PTFE connector
}
Insert the sulfuric acid container in the holders.
}
Route the gas transfer line through the top right opening of the detection module.
}
Connect the gas transfer line to the sulfuric acid container:
– Insert the thin hose (3) into the sulfuric acid container.
– Connect the connector (4) to the sulfuric acid container.
– Connect the gas transfer line via the banjo bolt (5) on the connector.
NOTICE! Ensure the proper fit of the conical nipples!
}
Place the safety attachment on the sulfuric acid container and attach the safety clip.
}
Connect hose 20 with the safety attachment. Connect the hose to the measuring cell
later.
}
Connect the gas transfer line in the basic module:
– Route the free end of the gas transfer line through the top left opening on the
basic module.
– Connect the gas transfer line to the auto-protection valve assembly.
– Connect the heating cable of the gas transfer line to its socket.