multi EA 5000
Troubleshooting
105
Error
code
Error message/cause
Remedy
230
ABD: No connection
Communication error after ABD was detected dur-
ing initialization
Acknowledge message
Initialize
231
ABD: Runtime exceeded
Completion message for the ABD movement
takes too long
Check that the flame sensor is properly applied
and connected
Record log files
Inform customer service department
232
Flame sensor error
Flame sensor calibration failed
Record log files
Inform customer service department
250
LPG: No connection
Communication interrupted after LPG detection
during initialization
Acknowledge message
Initialize
251
LPG: Runtime exceeded
Communication error
Completion message at dosing end not received
Acknowledge message
Check argon gas inlet pressure
Initialize
252
LPG: Argon missing during dosing
No argon applied to LPG module
Check gas supply
Check gas inlet pressure
253
LPG: Incorrect sample volume
Dosed volume is not a whole multiple of the sam-
ple loop inserted
Adapt dosing volume to sample loop volume
260
Sample handling missing
No sampler module detected
Connect at least one sampler module
Initialize
270
Auto-injector syringe: No connection
(applies to Autoinjector, Autoinjector AI-EA)
No communication with auto-injector
Acknowledge message
Initialize
271
Auto-injector syringe: Runtime exceeded
(applies to Autoinjector, Autoinjector AI-EA)
Communication error
Completion message at dosing end not received
Acknowledge message
Check auto-injector
Initialize
272
Auto-injector syringe: Incorrect syringe size
Autoinjector: Dosing volume and syringe size dif-
ferent
Adapt the dosing volume and/or the syringe size
Autoinjector AI-EA cannot complete this order
Initialize
273
Auto-injector syringe: Syringe drawn incorrectly