multi EA 5000
Operation
37
6
Operation
6.1
General notes for measuring operations
Only use the correct sampling modules for the corresponding matrix and installation po-
sition of the furnace for sampling.
Furnace installation
position
Sample type
Sampling
Vertical:
Liquids
¡
Autoinjector
¡
Multi Matrix Sampler
MMS
Gas, non-pressurized
¡
GSS module
Gas, pressurized
¡
GSS/LPG combi module
¡
GSS module with
GSS adapterbox
LPG
¡
LPG module 2.0
¡
GSS/LPG combi module
Horizontal:
Solids
¡
Automatic Boat Drive
ABD
¡
ABD with MMS
Liquids
¡
ABD with manual dosing
syringe
¡
ABD with MMS
¡
Autoinjector
Gas, non-pressurized
¡
GSS module
Gas, pressurized
¡
GSS/LPG combi module
¡
GSS module with
GSS adapterbox
LPG
¡
LPG module 2.0
¡
GSS/LPG combi module
Observe the following during analysis:
¡
To add samples of flammable substances in horizontal operation, always use ABD
with a flame sensor in automatic or automatic plus mode or an Autoinjector.
¡
To add samples of flammable substances, do not use the ABD in parameter mode,
or only use it with program parameters designed and tested for it (risk of soot!).
¡
If the furnace is installed horizontally, solids can also be added with the MBD man-
ual sampling module. The MBD may not be used to add flammable substances (risk
of soot).
¡
Do not exceed the maximum permissible sample amounts (→ "Technical data for
the multi EA 5000 "
¡
Adjust the dosing rate to the sample matrix and observe the maximum dosing
speeds (→ "Technical data for the multi EA 5000 "
176).
¡
Standard solutions with organic solvents can change composition rapidly due to the
volatility. Therefore, ensure that the clear space above the liquid in the sample cup
is small when preparing and storing samples. Store the solutions in the refrigerator.
The boiling points of the materials used should also not differ by more than 50 °C.
¡
Adjust the sample volume to the expected concentration to remain within the mea-
suring range of the detector.