Troubleshooting
multi EA 5000
108
13.4 Displays in the window Status analyzer
13.4.1 Overview
In the Status analyzer window, information on the device status or information on indi-
vidual modules is displayed.
TN liquid vertical (1) - liquid
Status analyzer
multi EA 5000
Rack
Syringe
: 112 -
: 50 µL
C-NDIR
N-CLD-5000
S-UVFD-5000
MFC 1
MFC 2
MFC 3
Furnace temperature
Cl-POT
OK
0.15
300
0
100
1050 °C
1
2
3
4
5
6
Figure 65Status analyzer window
1 Window name and basic module
2 Active method and sample state
3 Sampling modules
4 Detectors with sensor values
5 Gas flow displays
6 Furnace temperature (horizontal
line – horizontal installation, vertical
line – vertical installation)
The displays in the Status analyzer window are color-coded. The colors have the follow-
ing meanings:
Color
Description
Black
Status of the corresponding component OK, device ready for measure-
ment
Gray
Detector is inactive
Green
Detector OK, device ready for measurement (OK)
Or
Detector busy, measurement can only be started once the routine is
complete (detector-specific)
Red
Component not ready for measurement:
¡
Run-in period not yet complete; wait for run-in time to complete
¡
Error: Troubleshooting, access the corresponding information for
the component via the System |Component test menu item.