
Hazardous Area
X-STREAM O
2
Transmitter
4-24
Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
D/A Trim with Xi
For systems with configuration 2, shown in Figure 4-10, use the handheld 375
Field Communicator to access the D/A trim procedure according to the
instructions that follow. Refer to the 375 Field Communicator Xi Menu Tree in
Figure 4-8.
1. From the DEVICE SETUP screen select menu item 2, DIAG/SERVICE,
to access the diagnostics and service menu options.
2. Select menu item 3, CALIBRATION, to access the calibration menu
options.
3. Select menu item 3, 4-20 mA SIG, to access the 4-20 mA SIGNAL
screen.
4. Select menu item 1, D/A Trim, to start the trim procedure.