Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
2-23
Hazardous Area
X-STREAM O
2
Transmitter
PNEUMATIC
INSTALLATION
Reference Air Package
After the X-STREAM is installed, connect the reference air set to the
X-STREAM unit. Refer to the schematic diagram in Figure 2-21 and the
mounting dimensions in Figure 2-22 for a locally assembled reference air
supply.
Instrument Air (Reference Air): 5 psi (34 kPa) minimum, 8 psi (54 kPa)
maximum at 0.5 scfh (0,25
/min) maximum; less than 40 parts per million total
hydrocarbons. Regulator outlet pressure should be set at 5 psi (34 kPa).
Reference air can be supplied by the reference air set or the optional SPS
4001B or IMPS 4000.
Figure 2-21. Plant Air Schematic Diagram