Hazardous Area
X-STREAM O
2
Transmitter
3-2
Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
Xi Configuration
(if equipped)
Refer to Figure 3-1 for the configuration of jumpers JP1 through JP8. The
jumper configuration for your I/O board depends on the X-STREAM system
design and system components used in your installation.
The setting of switch SW4 and the configuration of jumpers JP1 through JP8
must be verified on the I/O board in the Xi Enhanced Interface box. All four dip
switches on switch SW4 must be set to the OFF position, as shown.
Figure 3-1. I/O Board Jumper Configuration
SET TEST GAS VALUES
Use a Field Communicator or the optional Xi Enhanced Interface to set test
gas values for calibration.
A Hazardous Area X-STREAM Transmitter and Xi Enhanced Interface
shipped from the factory have test gas values for low and high set to 0.4%
and 8.0% respectively. This same process must be performed any time a
replacement Transmitter Board, IO Board or DR Board is installed.
Set Test Gas Values Using Field Communicator - Transmitter Only or
Xi Enhanced Interface
1. Use the 375 Field Communicator software to access the HART menu.
2. From the DEVICE SETUP menu, select DETAILED SETUP
3. From the DETAILED SETUP menu, select CAL SETUP.
4. From CAL SETUP, select Cal Gas 1. Enter the percent O2 used for the
low O2 test gas.
5. From CAL SETUP, select Cal Gas 2. Enter the percent O2 used for the
high O2 test gas.