Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
Hazardous Area
X-STREAM O
2
Transmitter
http://www.raihome.com
Section 3
Configuration of Hazardous Area
X-STREAM O
2
Transmitter
Verify Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Xi Configuration (if equipped) . . . . . . . . . . . . . . . . . . . . . . page 3-2
Set Test Gas Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-2
Alarm Relay Output Configuration . . . . . . . . . . . . . . . . . . page 3-3
Analog Output Configuration . . . . . . . . . . . . . . . . . . . . . . page 3-4
Autocalibration Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-5
Optional Advanced Features Inside the Xi . . . . . . . . . . . . page 3-5
Extended Process Temperature Range to 800°C . . . . . . . page 3-6
Stoichiometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-6
Programmable Reference . . . . . . . . . . . . . . . . . . . . . . . . . page 3-7
VERIFY INSTALLATION
Ensure the Hazardous Area X-STREAM O
2
Transmitter and Xi are installed
and wired correctly according to Section 2: Installation.
Transmitter
1. Refer to Figure 6-1. Remove right cover (12) to expose the transmitter
electronics board (8).
2. Check the electronics board wiring (Figure 2-11). Be sure the power
and 4-20 mA signal are properly connected and secure. To avoid a
shock hazard, the power terminal cover must be installed.
3. Install right cover (12).
4. Remove the left cover (18). Verify that the reference air and calibration
gas hoses (Figure 6-3) are properly connected. Then, reinstall the left
cover (18).
Install all protective equipment covers and safety ground leads before equipment startup.
Failure to install covers and ground leads could result in serious injury or death.
If external loop power is used, the power supply must be a safety extra low voltage (SELV)
type.