Hazardous Area
X-STREAM O
2
Transmitter
1-10
Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
SPECIFICATIONS
Measurement Specifications
Net O
2
Range:
0 to 50% O
2
user scalable
-2 to 50% O
2
user scalable with stoichiometer
Accuracy in Oxidizing Conditions:
±0.75% of reading or 0.05% O
2
, whichever is greater
Lowest Detectable Limit:
0.01% O
2
Signal Stability:
±0.03% O
2
Process Temperature Effect:
less than 0.05% O
2
from 100° to 700°C
(212° to 1292°F)
System Speed of Response to
Calibration Gas:
Initial response in less than 3 seconds
T in less than 8 seconds
Response to process gas changes will vary depending
on velocity and particulate loading of the diffuser
Calibration Validity:
Presentation of calibration gases matches the normal
process to within ±0.02% O
2
Accuracy in Reducing Conditions:
±10% of reading or 0.1% O
2
System Response in Reducing
Conditions:
going from oxidizing to reducing -T
90
in 120 seconds
going from reducing to oxidizing -T
90
in 30 seconds
Ambient Temperature Effect on
Transmitter 4-20 mA Signal:
less than 0.005% O
2
per degree Celsius
Ambient Temperature Effect on Xi
4-20 mA Signal:
less than 0.0025% O
2
per degree Celsius
Environmental Specifications
Transmitter Probe:
Process-wetted materials are 316L or 304 Stainless
Process Temperature Limits:
0° to 800°C (32° to 1472°F)*
*Reduced cell life can be expected if operated
continuously at temperatures above 705°C (1300°F)
[optional bypass and jacket accessories permit
operation to 1050°C (1922°F)]
Transmitter Electronics Housing: Low copper aluminum Type 4X/IP66, with reference air
exhaust port piped to clean, dry area
Ambient Temperature Limits:
-40° to 80°C (-40° to 176°F), Transmitter
-40° to 85°C (-40° to 185°F) as measured by electronics
-40° to 100°C (-40° to 212°F), DR Probe (Model
XS-O2XP only)
Process Mounting Temperature:
200°C (392°F) Maximum
Xi Enhanced Interface:
Type 4X/IP66, Polycarbonate Material
Ambient Temperature Limits:
-20° to 50°C (-4° to 122°F)
-20° to 70°C (-4° to 158°F) as measured by electronics
Xi LCD display:
Ambient Temperature Limits
-20° to 55°C (-4° to 131°F)
Agency Certifications
Model XS-O2XP (without process end flame arrestor)
Electronics Options
11, 12, 13
Electronics Options
21, 22, 23
Special Condition of Use
200°C (392°F) maximum flange mounting temperature
Model XS-O2XPFA (with process end flame arrestor)
Electronics Options
01, 02, 03
∞
Standards:
EN 60079-0:2004
EN 60079-1:2003
Standards:
EN 60079-0:2006
EN 60079-1:2007