Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
3-5
Hazardous Area
X-STREAM O
2
Transmitter
Configuring the Analog Output with the Xi Enhanced Interface
Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select ANALOG OUTPUT.
5. From ANALOG OUTPUT, set the following parameters:
O2 LRV
- O
2
value at the lower analog output value (0 mA or 4 mA)
O2 URV
- O
2
value at the upper analog output value (20 mA)
AO Range
- Range of the analog output (0-20 mA or 4-20 mA)
Signal Alarm Level
- O
2
alarm level (3.5 mA or 21.1 mA)
AUTOCALIBRATION
SETUP
If autocalibration is desired, the Xi Enhanced Interface must be used with
either an SPS 4001B or IMPS 4000. The Xi Enhanced Interface must be
properly configured before autocalibration can take place. Refer to the
applicable SPS 4001B or IMPS 4000 instruction manual for details on
performing autocalibration. Refer to Section 4, Startup and Operation for
details on manual calibration procedures.
An Xi Enhanced Interface is shipped from the factory without autocalibration
configured. This same process must be performed any time a replacement
IO Board is installed.
Configuring Autocalibration with the Field Communicator
1. Use the 375 Field Communicator software to access the HART menu.
2. From the DEVICE SETUP menu, select DETAILED SETUP.
3. From the DETAILED SETUP menu, select CAL SETUP.
4. From CAL SETUP, select Auto Cal.
5. Press the RIGHT arrow key to change the state from NO to YES.
Configuring Autocalibration with the Xi Enhanced Interface
Keypad/Display
1. Press the MENU button once.
2. From the main menu, select PROBE 1.
3. From PROBE 1, select DETAILED SETUP.
4. From the DETAILED SETUP menu, select CAL SETUP.
5. From CAL SETUP, select Auto Cal.
6. Press the RIGHT arrow key to change the state from NO to YES.
OPTIONAL ADVANCED
FEATURES INSIDE
THE XI
Advanced features available inside the Xi Enhanced Interface are typically
purchased with the Hazardous Area X-STREAM Transmitter. However, these
advanced features are also available for field retrofit.
A Xi Enhanced Interface is shipped from the factory with the optional
enhanced software features enabled based on the configuration.