Before putting fresh reagent in the carboy, discard any small amount of remaining
reagent. To prepare the reagent refer to the procedure in
. See
ordering information.
10.3.2
Sample and reagent tubing
Periodically inspect sample and reagent tubing for cracks and leaks. Replace tubing if it is
damaged.
After a period of time, the sample tubing may become plugged with suspended matter.
The tubing is flexible and difficult to clean mechanically. Plugged sample tubing is best
replaced.
Replacement tubing kits are available. See
.
.
Replacing Reagent Tubing
Figure 10-3:
To replace reagent tubing:
Procedure
1.
Turn off sample and reagent pumps.
2.
Luer fittings connect the reagent tubing to the pump. Disconnect the tubing by
turning the fitting in the direction of the arrows shown in
3.
Disconnect the other end of the suction tubing from the barb on the reagent inlet
fitting in the bottom of the enclosure. Disconnect the other end of the discharge
tubing from the reagent injection tee.
4.
Install the replacement tubing.
Maintenance
Instruction Manual
75