3.
Press MENU.
The main Menu screen appears. The cursor is on
Calibrate
.
4.
Press ENTER.
5.
Choose the sensor you wish to calibrate.
Sensor 2
appears only if you are using a dual input transmitter.
6.
Choose
Temperature
.
7.
Change the display to match the temperature read from the calibrated
thermometer. Press ENTER.
If the present temperature is more than 5 °C different from the value entered, an
error message appears.
8.
To force the transmitter to accept the calibration, choose
Yes
. To repeat the
calibration, choose
No
.
For troubleshooting assistance, see
.
9.
To return to the main display, press MENU and then EXIT.
Calibration
62
Rosemount TCL