Procedure
1.
Connect the sample line to the sample conditioning system. Use 1/4 in. OD hard
plastic or stainless steel tubing. If dechlorinated water is being measured, provide a
way for occasionally substituting a chlorinated water sample for the dechlorinated
sample.
Chlorinated water is needed to calibrate the sensor and to check its response.
2.
If a grab sample is not already available, install one in the process piping. Choose a
point as close as possible to the sample line supplying the TCL.
Be sure that opening the sample valve does not appreciably alter the flow of sample
to the instrument.
3.
Connect the drain to a length of 3/4 in. ID flexible plastic tubing.
Important
The sample must drain to open atmosphere.
4.
Find the reagent tubing and fitting in the plastic bag taped to the inside of the
enclosure door. Screw the reagent fitting onto the bulkhead fitting at the bottom
left of the enclosure. Pass the reagent tubing through the hole in the carboy cap. Be
sure the plastic weight will be inside the carboy when the cap is in place. Attach the
reagent tubing to the barbed connector.
.
5.
Place the blue plastic carboy beneath the enclosure. Screw the cap and tubing
assembly on the carboy.
3.2.4
Install the sensor
Complete the following steps to install the Rosemount 499ACL-02 Sensor in the TCL
system.
1.
From inside the sample conditioning enclosure, thread the sensor cable or VP cable
through the gland on the upper left side.
Leave about one foot of cable inside the enclosure.
2.
Wire the cable to the transmitter.
Refer to
3.
Remove the nut and adapter from the flow cell.
4.
Slip the nut over the end of the sensor.
5.
Thread the adapter onto the sensor. Hand-tighten only.
6.
If you are using a VP cable, connect the cable to the sensor.
The connector and receptacle are keyed to ensure proper mating.
Installation
16
Rosemount TCL
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...