1.
Choose temperature units.
2.
Choose automatic or manual temperature correction for membrane permeability
(chlorine sensor).
3.
Enter a temperature for manual temperature compensation.
7.6.2
Definitions - chlorine
Automatic
temperature
correction
The total chlorine sensor is a membrane-covered amperometric
sensor. The permeability of the membrane is a function of
temperature. As temperature increases, membrane permeability
increases. Thus, an increase in temperature will cause the sensor
current and the transmitter reading to increase even though the
chlorine level remained constant. In automatic temperature
correction, the transmitter uses the temperature measured by the
sensor for the correction.
Manual
temperature
correction
In manual temperature correction, the transmitter uses the
temperature you enter for correction. It does not use the actual
process temperature. Do not use manual temperature correction
unless the measurement and calibration temperatures differ by no
more than about 2 °C. Manual temperature correction is useful if the
sensor temperature element has failed and a replacement sensor is not
available.
7.6.3
Procedure - configuring temperature related settings
Complete the following steps to set the temperature units and to select automatic or
manual temperature correction.
1.
Press MENU.
The main Menu screen appears.
2.
Move the cursor to
Program
and press ENTER.
3.
Choose
Temperature
.
The screen summarizes the present sensor configuration. There are between two
and five items.
Units
and
S1 Temp Comp
always appear. If manual temperature
compensation was selected, the manual temperature value also appears. If you have
Programming the transmitter
54
Rosemount TCL
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...