4.
Press ENTER.
The Output Range screen appears. The screen shows the present values assigned to
output 1 (
O1
) and output 2 (
O2
). The screen also shows which sensors the outputs
are assigned to.
S1
is sensor 1, and
S2
is sensor 2. S2 appears only if you have a dual
input Rosemount 1056 transmitter. The assignments shown are the defaults for a
single channel chlorine transmitter. Outputs are freely assignable under the
Configure menu.
5.
For practice, change the 20 mA settings for output 1 to 8.5 ppm.
a. Move the cursor to the
O1 S1 20 mA: 10.00
line and press ENTER.
The screen below appears.
b. Use the navigation keys to change
10.00
to
8.5
ppm. Use Left and Right to
move from digit to digit. Use Up and Down to increase or decrease the numeral.
c. To move the decimal point, press Left or Right until the decimal point is
highlighted. Press Up to move the decimal point to the right. Press Down to
move to the left.
d. Press ENTER to store the setting.
The display returns to the summary screen. Note that the 20 mA setting for output 1
has changed to
8.50 ppm
.
Display and operation
Instruction Manual
33
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...