Note that the discharge tubing is longer than the suction tubing.
10.3.3
Replacing sample tubing
To replace sample tubing:
Sample tubingis shown in
Replacing sample tubing
Figure 10-4:
Procedure
1.
Turn off the sample and reagent pumps.
2.
Luer fittings connect the sample tubing to the pump. Disconnect the tubing by
turning the fitting in the direction of the arrows shown in
3.
Disconnect the other end of the sample pump suction tubing from the overflow
sampler. Pull the reagent injection tubing off the reagent injection tee.
4.
Disconnect the other end of the sample pump discharge tubing from the flow cell.
Pull the air injection tubing off the air injection tee.
5.
Disconnect the sample inlet and drain tubing.
6.
Install the replacement sample pump suction and discharge tubing assemblies.
Maintenance
76
Rosemount TCL