7
Programming the transmitter
7.1
General
This section describes how to make the following program settings using the local keypad.
1.
Configure and assign values to the analog current outputs.
2.
Configure and assign values to the alarm relays (if the alarm board is installed).
3.
Choose the type of chlorine measurement being made. This step is necessary
because the transmitter used with the TCL can measure forms of chlorine other than
total chlorine.
4.
Choose temperature units or manual temperature correction.
5.
Set two levels of security codes.
6.
Reset the transmitter to factory default settings.
7.2
Default settings
The transmitter leaves the factory with the default settings for total chlorine shown in
. You can change the settings to any value shown in the column labeled Choices. If
you have a dual input Rosemount 1056 Transmitter, refer to the Rosemount 1056
Transmitter Instruction Manual for information about the default settings for the second
input.
Default Settings
Table 7-1:
Item
Choices
Default
Outputs
1. Assignments
a. output 1
chlorine, temp
chlorine
b. output 2
chlorine, temperature
temperature
2. Range
0-20 or 4-20 mA
4-20 mA
3. 0 or 4 mA setting
a. chlorine
-9999 to +9999
0
b. temperature
-999.9 to +999.9
0
4. 20 mA setting
a. chlorine
-9999 to +9999
10
b. temperature
-999.9 to +999.9
0
5. Fault current (fixed)
0.00 to 22.0 mA
22.0 mA
6. Dampening
0 to 999 sec
0 sec
Programming the transmitter
Instruction Manual
39
Summary of Contents for Rosemount TCL
Page 4: ......
Page 8: ...Contents iv Rosemount TCL ...
Page 16: ...Description and specifications 8 Rosemount TCL ...
Page 18: ...Principles of operation 10 Rosemount TCL ...
Page 26: ...Installation 18 Rosemount TCL ...
Page 34: ...Wiring 26 Rosemount TCL ...
Page 46: ...Display and operation 38 Rosemount TCL ...
Page 66: ...Programming the transmitter 58 Rosemount TCL ...
Page 68: ...Digital communications 60 Rosemount TCL ...
Page 106: ...Troubleshooting 98 Rosemount TCL ...